4- 78 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO:LLIMit
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
RANGe:AUTO:LLIMit
Specifies the lower limit for the automatic measurement ranging operation, and sets
the minimum measurement range which provides the best resolution to measure the
specified value.
If the minimum measurement range is the same as the maximum measurement
range, the measurement is performed by using this range.
An error occurs if the minimum range is greater than the maximum range.
Syntax :SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO:LLIMit
range
:SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO:LLIMit?
[range]
For <CURRent[:DC]|RESistance|VOLTage[:DC]>, specify CURRent[:DC] for
current measurement, RESistance for resistance measurement, or VOLTage[:DC]
for voltage measurement.
Parameter range value|MINimum|MAXimum|DEFault. Parameter data type is
NRf+. Query does not support range=value.
value for current measurement: See Table 2-8.
value for voltage measurement: See Table 2-7.
value for resistance measurement: See Table 2-9. This is available for the resistance
measurements set to the AUTO mode which is selected by the
:SENSe:RESistance:MODE command.
Query response range <newline>
range returns the present setting of the minimum measurement range for the auto
range. If a parameter is specified, range returns the value assigned to DEF, MIN, or
MAX. Response data type is NR3.
Example :SENS:CURR:RANG:AUTO:LLIM 1E-6
:SENS2:CURR:DC:RANG:AUTO:LLIM?