5- 16 Agilent B2900 SCPI Command Reference, Edition 2
Error Messages
B2900 Specific Error
Error 324 Emergency; SMU power supply was turned off; Channel n
SMU power supply was turned off because emergency status was detected in
channel n. All channels were disabled. Execute the *TST? command.
Error 331 Emergency; Interlock open detected
Interlock open was detected. All channel output is changed to 0 V and the all output
switch is opened. Execute the *TST? command. Do not open interlock circuit while
SMU is in high voltage state.
Error 341 Emergency; Fan speed is too slow
Too slow fan speed status was detected. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 342 Emergency; Fan speed is too fast
Too fast fan speed status was detected. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 351 Emergency; Internal communication failure detected by
SMU; Channel n
Internal communication failure was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 352 Emergency; Watchdog timer expired; Channel n
Watchdog timer expired status was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 353 Emergency; F-COM CPLD reset detected; Channel n
F-COM CPLD reset status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 354 Emergency; VADC data was lost; Channel n
Channel n voltage ADC data was lost. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 355 Emergency; IADC data was lost; Channel n
Channel n current ADC data was lost. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.