Vedere la pagina 283
4- 162 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:TOUTput:SIGNal
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:TOUTput:S
IGNal
Selects the trigger output for the status change between the arm layer and the trigger
layer. Multiple trigger output ports can be set.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:TOUTput:SIGNal output{,output}
:TRIGger[c]<:ACQuire|:TRANsient>:TOUTput:SIGNal?
For <:ACQuire|:TRANsient|[:ALL]> and <:ACQuire|:TRANsient>, specify
:ACQuire for measurement, :TRANsient for source output, or :ALL for both device
actions.
Parameter output Trigger output port.
EXT1
(default)
|EXT2|EXT3|EXT4|EXT5|
EXT6|EXT7|EXT8|EXT9|EXT10|EXT11|EXT12|EXT13|EXT14|
LAN|INT1|INT2
. Parameter data type is CPD.
output=INT1 or INT2 selects the internal bus 1 or 2, respectively.
output=LAN selects a LAN port.
output=EXTn selects the GPIO pin n, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting, INT1, INT2, LAN, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :TRIG:TOUT:SIGN EXT3
:TRIG2:TRAN:TOUT:SIGN?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:TOUTput[:
STATe]
Enables or disables the trigger output for the status change between the arm layer
and the trigger layer.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:TOUTput[:STATe] mode
:TRIGger[c]<:ACQuire|:TRANsient>:TOUTput[:STATe]?