4- 34 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:FORMat:DIGital
format=ASCii specifies the ASCII format (default).
format=REAL,32 specifies the IEEE-754 single precision format. 4-byte data.
format=REAL,64 specifies the IEEE-754 double precision format. 8-byte data.
Query response format <newline>
format returns ASC, REAL,32, or REAL,64. Response data type is CRD.
Example :FORM REAL32
:FORM?
:FORMat:DIGital
Sets the response format of the bit pattern defined by the following commands.
• :CALCulate:CLIMits:<FAIL|PASS>:DIGital[:DATA]
• :CALCulate:LIMit:COMPliance:DIGital[:DATA]
• :CALCulate:LIMit:<LOWer|UPPer>:DIGital[:DATA]
• :CALCulate:LIMit:PASS:DIGital[:DATA]
Syntax :FORMat:DIGital format
:FORMat:DIGital?
Parameter format Response format. ASCii (decimal, default)|
BINary|OCTal|HEXadecimal. Parameter data type is CPD.
Query response format <newline>
format returns ASC, BIN, OCT, or HEX. Response data type is CRD.
Example :FORM:DIG BIN
:FORM:DIG?
:FORMat:ELEMents:CALCulate
Specifies the elements included in the calculation result data returned by the
:CALCulate:DATA?, :CALCulate:DATA:LATest?, :CALCulate:MATH:DATA?,
:CALCulate:MATH:DATA:LATest?, or :TRACe:DATA? command.