Agilent Technologies B2902 Manuale di Servizio Pagina 65

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 326
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 64
Agilent B2900 SCPI Command Reference, Edition 2 2- 25
Subsystem Command Summary
Setting Source/Measure Unit
Table 2-6 Current Output Range
Range
value
Setting resolution
DC output current or
pulse peak/base
current
a
b
Maximum voltage
Pulse width t
c
B2901A
B2902A
B2911A
B2912A
DC
output
Pulsed
output
10 nA
d
10 fA 0 |I| 10.5 nA 210 V 210 V 50 s t t
max
100 nA 1 pA 100 fA 0 |I| 105 nA
1 A 10 pA 1 pA 0 |I| 1.05 A
10 A 100 pA 10 pA 0 |I| 10.5 A
100 A 1 nA 100 pA 0 |I| 105 A
1 mA 10 nA 1 nA 0 |I| 1.05 mA
10 mA 100 nA 10 nA 0 |I| 10.5 mA
100 mA 1 A 100 nA 0 |I| 105 mA
1 A 10 A 1 A 0 |I| 105 mA
105 mA < |I| 1.05 A 21 V 21 V
0 |I| 1.05 A
200 V 50 s t 2.5 ms
180 V 50 s t 10 ms
1.5 A 0 |I| 105 mA 210 V 210 V 50 s t t
max
105 mA < |I| 1.515 A 21 V 21 V
0 |I| 1.515 A
200 V 50 s t 2.5 ms
0 |I| 1.05 A
180 V 50 s t 10 ms
3 A 100 A 10 A 0 |I| 105 mA 210 V 210 V 50 s t t
max
105 mA < |I| 1.515 A 21 V 21 V
1.515 A < |I| 3.03 A 6 V 6 V
10 A 0 |I| 10.5 A
6 V 50 s t 1 ms
0 |I| 1.515 A
200 V 50 s t 2.5 ms
0 |I| 1.05 A
180 V 50 s t 10 ms
Vedere la pagina 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 325 326

Commenti su questo manuale