Agilent B2900 SCPI Command Reference, Edition 2 4- 107
Subsystem Commands
[:SOURce]:PULSe:WIDTh
Parameter delay Delay time. value (0.0 to 99999.9, in seconds)|MINimum|
MAXimum|DEFault (default is 0). Parameter data type is
NRf+. Query does not support delay=value.
Query response delay <newline>
delay returns the present setting. If a parameter is specified, delay returns the value
assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :PULS:DEL 1E-3
:SOUR2:PULS:DEL?
[:SOURce]:PULSe:WIDTh
Sets the pulse width for the specified channel. The pulse width is the time from
starting the pulse peak output (or starting the pulse level transition) to the end of the
pulse peak output. However, it is strictly defined as the time from 10 % of peak level
at the leading edge to 90 % of peak level at the trailing edge.
Syntax [:SOURce[c]]:PULSe:WIDTh width
[:SOURce[c]]:PULSe:WIDTh? [width]
Parameter width Pulse width. value (5E-5 to 100000 seconds, in 1E-6
resolution)|MINimum| MAXimum|DEFault (default is 5E-5).
Parameter data type is NRf+. Query does not support
width=value.
Minimum time for the pulse base output is also 50 s. And the
minimum pulse period is 100 s.
Query response width <newline>
width returns the present setting. If a parameter is specified, width returns the value
assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :PULS:WIDT 2E-2
:SOUR2:PULS:WIDT?
[:SOURce]:SWEep:DIRection
Sets the sweep direction, UP or DOWN, for the specified channel.