Agilent B2900 SCPI Command Reference, Edition 2 4- 125
Subsystem Commands
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTa
ge>:PTRansition
filter returns the present setting of the negative transition filter in the specified
register. Response data type is NR1 (decimal) or NDN (binary, octal, or
hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:QUES:CURR:NTR 0
:STAT:QUES:TEMP:NTR?
:STATus:QUEStionable:<CALibration|CURRent|TEM
Perature|TEST|VOLTage>:PTRansition
Sets the positive transition filter in the questionable calibration, current,
temperature, test, or voltage status register. If you set a bit of the filter, a 0-to-1
transition of its register bit sets the corresponding bit of the event register.
Syntax :STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:P
TRansition filter
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:P
TRansition?
For <CALibration|CURRent|TEMPerature|TEST|VOLTage>, specify CALibration
for the questionable calibration status condition register, CURRent for the
questionable current status condition register, TEMPerature for the questionable
temperature status condition register, TEST for the questionable self-test status
condition register, or VOLTage for the questionable voltage status condition register.
Parameter filter Positive transition filter. 0 to 65535 (decimal). Default is 32767.
Parameter data type is NR1 or NDN.
filter is the sum of the binary-weighted values for the set bits.
Query response filter <newline>
filter returns the present setting of the positive transition filter in the specified
register. Response data type is NR1 (decimal) or NDN (binary, octal, or
hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:QUES:CURR:PTR 32767
:STAT:QUES:TEMP:PTR?
Commenti su questo manuale