Agilent B2900 SCPI Command Reference, Edition 2 4- 133
Subsystem Commands
:SYSTem:COMMunicate:LAN:TELNet:WMESsage
Parameter prompt Command prompt. Up to 15 characters. Parameter data type is
SPD.
Query response prompt <newline>
prompt returns the command prompt. Response data type is SRD.
Example :SYST:COMM:LAN:TELN:PROM “A-B2911A-00001> ”
:SYST:COMM:LAN:TELN:PROM?
:SYSTem:COMMunicate:LAN:TELNet:WMESsage
Sets the welcome message displayed during a Telnet session when starting
communication with the instrument. This command setting is not changed by power
off or the *RST command.
The instrument uses LAN port 5024 for SCPI Telnet sessions, and 5025 for SCPI
Socket sessions.
Syntax :SYSTem:COMMunicate:LAN:TELNet:WMESsage message
:SYSTem:COMMunicate:LAN:TELNet:WMESsage?
Parameter message Welcome message. Up to 63 characters. Parameter data type is
SPD.
Query response message <newline>
message returns the welcome message. Response data type is SRD.
Example :SYST:COMM:LAN:TELN:WMES “Welcome to A-B2911A-00001.”
:SYST:COMM:LAN:TELN:WMES?
:SYSTem:COMMunicate:LAN:UPDate
Disconnects all active LAN and Web Interface connections, updates the LAN setup,
and restarts the LAN interface with the new setup. The new setup may change the IP
address of the instrument.
Syntax :SYSTem:COMMunicate:LAN:UPDate
Example :SYST:COMM:LAN:UPD