Agilent B2900 SCPI Command Reference, Edition 2 4- 7
Subsystem Commands
:CALCulate:DATA?
result returns IMM or END. Response data type is CRD.
Example :CALC:CLIM:UPD END
:CALC2:CLIM:UPD?
:CALCulate:DATA?
Returns limit test data. Elements of the returned data are specified by the
:FORMat:ELEMents:CALCulate command. The limit test data can be expressed by
the following formula.
limit test data = input data - null offset
input data: Data specified by :CALCulate:FEED
null offset: Data set by :CALCulate:OFFSet or :CALCulate:OFFSet:ACQuire
If the null offset function is disabled by the :CALCulate:OFFSet:STATe command,
null offset=0.
Syntax :CALCulate[c]:DATA? [offset[, size]]
Parameter offset Indicates the beginning of the data received. n|CURRent|STARt
(default). Parameter data type is NR1 or CPD.
offset=n specifies the n+1th data. n is an integer, 0 to maximum
(depends on the buffer state).
offset=CURR specifies the present data position.
offset=STAR specifies the top of the data buffer. Same as
offset=0.
size Number of data to be received. 1 to maximum (depends on the
buffer state). Parameter data type is NR1. If this parameter is
not specified, all data from offset is returned.
Query response data <newline>
Response data type is NR3. See “Data Output Format” on page 1-12.
Example :CALC2:DATA? 0,10