B2900 Specific Error
Error 713 ProgramMemory; Cannot set multiple INIT commands in
program line
A program line cannot contain multiple INIT commands.
Error 714 ProgramMemory; Invalid character in program line
Program line contains invalid character. Use appropriate characters.
Error 715 ProgramMemory; Invalid character in program name
Appropriate program name must be specified.
Error 716 ProgramMemory; Program count overflow
Program memory cannot save the program. Delete dispensable program.
Error 801 Calculate; Expression list full
Cannot save the expression. Delete dispensable expression.
Error 802 Calculate; Expression cannot be deleted
Cannot delete the specified expression. Specify erasable expression.
Error 803 Calculate; Missmatched parenthesis
Number of open and close parentheses must be the same.
Error 804 Calculate; Not a number of data handle
Expression contains invalid floating point number or symbol. Enter appropriate
expression. Available symbols are VOLT, CURR, RES, TIME, and SOUR.
Error 805 Calculate; Mismatched brackets
Number of open and close brackets must be the same.
Error 806 Calculate; Entire expression not parsed
Expression is not correct. Enter appropriate expression.
Error 807 Calculate; Not an operator or number
Expression contains not an operator or not a number. Enter appropriate expression.
Error 811 Calculate; Error parsing value