4- 114 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:WAIT[:STATe]
[:SOURce]:WAIT[:STATe]
Enables or disables the source wait time for the specified channel. This wait time is
defined as the time the source channel cannot change the output after the start of a
DC output or the trailing edge of a pulse.
Syntax [:SOURce[c]]:WAIT[:STATe] mode
[:SOURce[c]]:WAIT[:STATe]?
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the source wait time. The wait time is set to 0.
mode=1 or ON enables the source wait time given by the following formula.
[:SOURce]:WAIT:AUTO ON|1 condition:
wait time = gain initial wait time + offset
[:SOURce]:WAIT:AUTO OFF|0 condition:
wait time = offset
The initial wait time is automatically set by the instrument and cannot be changed.
gain and offset are set by the [:SOURce]:WAIT:GAIN and
[:SOURce]:WAIT:OFFSet commands respectively.
Query response mode <newline>
mode is 0 or 1, and indicates that the source wait time is disabled or enabled,
respectively. Response data type is NR1.
Example :WAIT 0