4- 92 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:<CURRent|VOLTage>:<CENTer|SPAN>
SOURce Subsystem
For the numeric suffixes [c] and [n], see “Numeric Suffix” on page 1-8.
[:SOURce]:<CURRent|VOLTage>:<CENTer|SPAN>
Sets the center or span value of the current or voltage sweep output.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:<CENTer|SPAN> data
[:SOURce[c]]:<CURRent|VOLTage>:<CENTer|SPAN>? [data]
For <CURRent|VOLTage>, specify CURRent for current output, or VOLTage for
voltage output.
For <CENTer|SPAN>, specify CENTer for the sweep center value, or SPAN for the
sweep span value.
Parameter data Sweep center or span value. value (see “Source Output Ranges”
on page 2-24) |MINimum|MAXimum|DEFault (default is 0.0).
Parameter data type is NRf+. Query does not support
data=value.
The center and span values can be expressed by the following formula, using the
start and stop values set by the
[:SOURce]:<CURRent|VOLTage>:<STARt|STOP>
command. So the last command setting is effective for these sweep parameters.
center = (start + stop)/2
span = stop - start
Query response data <newline>
data returns the present setting. If a parameter is specified, data returns the value
assigned to DEF, MIN, or MAX. Response data type is NR3.
Example :CURR:CENT 1E-3
:SOUR2:VOLT:SPAN?
[:SOURce]:<CURRent|VOLTage>[:LEVel][:IMMediate
][:AMPLitude]
Changes the output level of the specified source channel immediately.