Agilent B2900 SCPI Command Reference, Edition 2 4- 9
Subsystem Commands
:CALCulate:DIGital:<BUSY|EOT|SOT>
Query response pin <newline>
pin returns NONE or a comma separated EXT string. Response data type is CRD.
Example :CALC:DIG:BIT EXT9,EXT10,EXT11
:CALC2:DIG:BIT?
:CALCulate:DIGital:<BUSY|EOT|SOT>
Assigns the GPIO pin for the BUSY (busy), EOT (end of test), or SOT (start of test)
signal line for the composite limit test.
Syntax :CALCulate[c]:DIGital:<BUSY|EOT|SOT> pin
:CALCulate[c]:DIGital:<BUSY|EOT|SOT>?
For <BUSY|EOT|SOT>, specify BUSY for assigning the busy line, EOT for
assigning the end of test line, or SOT for assigning the start of test line.
Parameter pin EXTn|NONE (default). Parameter data type is CPD. EXTn
specifies a GPIO pin, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
To assign the GPIO pin, pin must be a EXT string like EXTn. For example,
pin=EXT14.
pin=NONE does not assign the GPIO pin.
Query response pin <newline>
pin returns NONE or a EXT string. Response data type is CRD.
Example :CALC:DIG:EOT EXT14
:CALC2:DIG:SOT?
:CALCulate:FEED
Specifies the input data value used for calculating the limit test data. The limit test is
a pass/fail judgement performed during a composite limit test. The limit test is
performed at the “Pass?” step in Figures
2-10 and 2-11. The limit test data is
returned by the :CALCulate:DATA? or :CALCulate:DATA:LATest? command.
Syntax :CALCulate[c]:FEED type