Agilent B2900 SCPI Command Reference, Edition 2 4- 143
Subsystem Commands
:SYSTem:TIME:TIMer:COUNt?
minute Minute. Integer from 0 to 59. Parameter data type is NR1.
second Second. Integer from 0 to 59. Parameter data type is NR1.
Query response response <newline>
response returns hour, minute, second. Each value is separated by a comma.
Response data type is NR1.
Example :SYST:TIME 23,59,59
:SYSTem:TIME:TIMer:COUNt?
Returns the present count of the timer.
Syntax :SYSTem:TIME:TIMer:COUNt?
Query response response <newline>
response returns the present timer count. Response data type is NR3.
Example :SYST:TIME:TIM:COUN?
:SYSTem:TIME:TIMer:COUNt:RESet:AUTO
Enables or disables the automatic reset function of the timer. If this function is
enabled, the timer count is reset when the initiate action occurs.
Syntax :SYSTem:TIME:TIMer:COUNt:RESet:AUTO mode
:SYSTem:TIME:TIMer:COUNt:RESet:AUTO?
Parameter mode Automatic reset function on or off. 0|OFF|1|ON (default).
Parameter data type is boolean.
mode=1 or ON enables the automatic reset function.
mode=0 or OFF disables the automatic reset function.
Query response mode <newline>
mode returns 0 or 1, and indicates that the automatic reset function is off or on,
respectively. Response data type is NR1.
Example :SYST:TIME:TIM:COUN:RES:AUTO 0