Agilent B2900 SCPI Command Reference, Edition 2 4- 111
Subsystem Commands
[:SOURce]:TOUTput:SIGNal
mode=DOUBle sets the sweep mode to double sweep. Double sweep performs the
sweep from start to stop to start.
Query response mode <newline>
mode returns SING or DOUB, and indicates that the sweep mode is single or
double, respectively. Response data type is CRD.
Example :SWE:STA DOUB
:SOUR2:SWE:STA?
[:SOURce]:TOUTput:SIGNal
Selects the trigger output for the status change between the trigger layer and the
transient device action. Multiple trigger output ports can be set.
Syntax [:SOURce[c]]:TOUTput:SIGNal output{,output}
[:SOURce[c]]:TOUTput:SIGNal?
Parameter output Trigger output port.
EXT1
(default)
|EXT2|EXT3|EXT4|EXT5|
EXT6|EXT7|EXT8|EXT9|EXT10|EXT11|EXT12|EXT13|EXT14|
LAN|INT1|INT2
. Parameter data type is CPD.
output=INT1 or INT2 selects the internal bus 1 or 2, respectively.
output=LAN selects a LAN port.
output=EXTn selects the GPIO pin n, which is an output port of the Digital I/O
D-sub connector on the rear panel. n=1 to 14.
Query response response <newline>
response returns the present setting, INT1, INT2, LAN, or EXT1 through EXT14.
Response data type is CRD. Multiple responses are separated by a comma.
Example :TOUT:SIGN EXT3
:SOUR2:TOUT:SIGN?
[:SOURce]:TOUTput[:STATe]
Enables or disables the trigger output for the status change between the trigger layer
and the transient device action.