Agilent Technologies B2902 Manuale di Servizio Pagina 216

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 326
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 215
4- 94 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:<CURRent|VOLTage>:MODE
Example :VOLT:TRIG 3
:SOUR2:CURR:LEV:TRIG:AMPL?
[:SOURce]:<CURRent|VOLTage>:MODE
Selects the source mode, fixed, list sweep, or sweep, of the specified source channel.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:MODE mode
[:SOURce[c]]:<CURRent|VOLTage>:MODE?
Parameter mode Source mode. SWEep|LIST|FIXed (default). Parameter data
type is CPD.
mode=FIX sets the constant current or voltage source.
mode=LIST sets the user-specified current or voltage list sweep source.
mode=SWEep sets the current or voltage sweep source.
Query response mode <newline>
mode returns FIX, LIST, or SWE. Response data type is CRD.
Example :VOLT:MODE SWE
:SOUR2:CURR:MODE?
[:SOURce]:<CURRent|VOLTage>:POINts
Sets the number of sweep steps for the current or voltage sweep output.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:POINts points
[:SOURce[c]]:<CURRent|VOLTage>:POINts? [points]
For <CURRent|VOLTage>, specify CURRent for current output, or VOLTage for
voltage output.
Parameter points Number of sweep steps. value (1 to 2500)|MINimum|
MAXimum|DEFault (default is 1). Parameter data type is
NRf+. Query does not support points=value.
The points value can be expressed by the following formula, using the step value set
by the
[:SOURce]:<CURRent|VOLTage>:STEP command and the span value set by
the [:SOURce]:<CURRent|VOLTage>:<CENTer|SPAN> command.
Vedere la pagina 215
1 2 ... 211 212 213 214 215 216 217 218 219 220 221 ... 325 326

Commenti su questo manuale

Nessun commento