Agilent B2900 SCPI Command Reference, Edition 2 4- 77
Subsystem Commands
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO
For <CURRent[:DC]|VOLTage[:DC]>, specify CURRent[:DC] for current
compliance, or VOLTage[:DC] for voltage compliance.
Query response status <newline>
status is 1 or 0, and indicates if the channel is in the compliance state or not.
Response data type is NR1.
Example :SENS:CURR:PROT:TRIP?
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
RANGe:AUTO
Enables or disables the automatic ranging function of the specified measurement
channel.
Syntax :SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO mode
:SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO?
For <CURRent[:DC]|RESistance|VOLTage[:DC]>, specify CURRent[:DC] for
current measurement, RESistance for resistance measurement, or VOLTage[:DC]
for voltage measurement.
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the automatic measurement ranging. If this function is
disabled, the measurement is performed by using the range set by the
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe[:UPPer] command.
mode=1 or ON enables the automatic measurement ranging. If this function is
enabled, the channel automatically sets the range which provides the best resolution
to perform the measurement.
If a range is manually selected, automatic ranging is disabled.
Query response mode <newline>
mode is 0 or 1, and indicates that the automatic measurement ranging is off or on,
respectively. Response data type is NR1.
Example :SENS:CURR:RANG:AUTO 0
:SENS2:CURR:DC:RANG:AUTO?