4- 86 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:REMote
Parameter function “CURRent[:DC]”|“VOLTage[:DC]”|“RESistance”. Parameter
data type is SPD.
function=“CURRent[:DC]” specifies the current measurement function.
function=“VOLTage[:DC]” specifies the voltage measurement function.
function=“RESistance” specifies the resistance measurement function.
Query response response <newline>
response returns 0 or 1, and indicates that the specified measurement function is
now disabled or enabled respectively. Response data type is NR1.
Example :SENS:FUNC:STAT? “CURR”
:SENSe:REMote
Enables or disables the remote sensing. Remote sensing must be enabled to use the
4-wire connection (Kelvin connection).
Syntax :SENSe[c]:REMote mode
:SENSe[c]:REMote?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables remote sensing.
mode=1 or ON enables remote sensing.
Query response mode <newline>
mode is 0 or 1, and indicates that the remote sensing is off or on, respectively.
Response data type is NR1.
Example :SENS:REM 1
:SENS2:REM?
:SENSe:RESistance:MODE
Selects the resistance measurement mode.
Syntax :SENSe[c]:RESistance:MODE mode
:SENSe[c]:RESistance:MODE?