2- 12 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
[:SOUR[c]]:PULS:WIDT width
[:SOUR[c]]:PULS:WIDT? [DEFault|M
INimum|MAXimum]
Sets the pulse width for the specified
channel. The pulse width is the time from
starting the pulse peak output (or starting
the pulse level transition) to the end of the
pulse peak output.See Figures
2-2 and
2-4.
width=DEFault|MINimum|MAXimum|
5E-5 to 100000 seconds, in 1E-6
resolution
Minimum time for the pulse base output
is also 50 s. And the minimum pulse
period is 100 s.
5E-5
seconds
[:SOUR[c]]:SWE:DIR direction
[:SOUR[c]]:SWE:DIR?
Sets the sweep direction, UP or DOWN,
for the specified channel.
direction=DOWN|UP
UP
[:SOUR[c]]:SWE:POIN points
[:SOUR[c]]:SWE:POIN? DEFault|MIN
imum|MAXimum
Sets the number of sweep steps for the
specified channel. This command setting
is effective for both current sweep and
voltage sweep.
points=DEFault|MINimum|MAXimum|
1
to 2500
The points value can be expressed by the
following formula, using the step value
set by the
[:SOUR[c]]:<CURR|VOLT>:STEP
command and the span value set by the
[:SOUR[c]]:<CURR|VOLT>:SPAN
command.
points = span/step + 1 (where step is not
0)
points=1 sets step=0.
1
Command Summary Reset setting