Agilent B2900 SCPI Command Reference, Edition 2 2- 43
Subsystem Command Summary
Reading Source/Measure Data
Table 2-14 MEASure Subsystem
Table 2-15 CALCulate Subsystem
Command Summary Reset setting
:MEAS? [chanlist]
Executes a spot measurement (one-shot
measurement) and returns the
measurement result data. Measurement
items can be specified by the
:FORM:ELEM:SENS command.
chanlist=(@1)|(@2)|(@1,2)|(@1:2)|
(@2,1)|(@2:1)
:MEAS:CURR[:DC]? [chanlist]
:MEAS:RES? [chanlist]
:MEAS:VOLT[:DC]? [chanlist]
Executes a spot measurement (one-shot
measurement) and returns the
measurement result data. Measurement
items can be specified by CURR, RES, or
VOLT.
chanlist=(@1)|(@2)|(@1,2)|(@1:2)|
(@2,1)|(@2:1)
Command Summary Reset setting
:CALC[c]:CLIM:CLE:AUTO mode
:CALC[c]:CLIM:CLE:AUTO?
Enables or disables the automatic clear
function of the composite limit test.
mode=0|OFF|1|ON
ON
:CALC[c]:CLIM:CLE:AUTO:DEL time
:CALC[c]:CLIM:CLE:AUTO:DEL?
Sets the delay time for the automatic clear
of the composite limit test.
time=MINimum|MAXimum|DEFault|
1E-5 to 60 seconds
Query does not support time=1E-5 to 60.
1E-4
:CALC[c]:CLIM:CLE[:IMM]
Clears the composite limit test results and
the GPIO lines immediately.