4- 146 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:TRACe:FEED
:TRACe:FEED
Specifies the data placed in the trace buffer. This command is effective when the
trace buffer control mode is set to NEV by the
:TRACe:FEED:CONTrol command.
Syntax :TRACe[c]:FEED type
:TRACe[c]:FEED?
Parameter type Data type. MATH|LIMit|SENSe (default). Parameter data type
is CPD.
type=SENS specifies the measurement result data, which contains all of the voltage
measurement data, current measurement data, resistance measurement data, time
data, status data, or source output setting data specified by the
:FORMat:ELEMents:SENSe command.
type=LIM specifies the limit test data. The data contains the limit test data, time
data, or status data specified by the
:FORMat:ELEMents:CALCulate command. See
:CALCulate:DATA? for the limit test data.
type=MATH specifies the calculation result data. The data contains the calculation
result, time data, or status data specified by the
:FORMat:ELEMents:CALCulate
command. See :CALCulate:MATH:DATA? for more information.
Query response type <newline>
type returns the present setting of data type, MATH, LIM, or SENS. Response data
type is CRD.
Example :TRAC:FEED MATH
:TRAC2:FEED?
:TRACe:FEED:CONTrol
Selects the trace buffer control.
Syntax :TRACe[c]:FEED:CONTrol mode
:TRACe[c]:FEED:CONTrol?
Parameter mode Trace buffer control mode. NEXT|NEVer (default). Parameter
data type is CPD.