Agilent B2900 SCPI Command Reference, Edition 2 4- 95
Subsystem Commands
[:SOURce]:<CURRent|VOLTage>:RANGe
points = span/step + 1 (where step is not 0)
points=1 sets step=0.
If points is changed, span works as a constant and step is changed. If step is
changed, span works as a constant and points is changed. If span is changed, points
works as a constant and step is changed.
The calculated points value is rounded down to an integer.
The sweep measurement is performed from the start value to the stop value given by
the following formula, even if the specified stop value does not satisfy it.
stop = start + step (points -1)
For the logarithmic sweep, the step value is ignored and is not used for the
calculation of sweep points.
Query response points <newline>
points returns the present setting. If a parameter is specified, points returns the value
assigned to DEF, MIN, or MAX. Response data type is NR1.
Example :CURR:POIN 51
:SOUR2:VOLT:POIN?
[:SOURce]:<CURRent|VOLTage>:RANGe
Sets the current or voltage output range of the specified source channel. This
command is effective when the automatic ranging function is off.
Syntax [:SOURce[c]]:<CURRent|VOLTage>:RANGe range
[:SOURce[c]]:<CURRent|VOLTage>:RANGe?
For <CURRent|VOLTage>, specify CURRent for current output, or VOLTage for
voltage output.
Parameter range value (see “Source Output Ranges” on page 2-24) |MINimum|
MAXimum|DEFault. Parameter data type is NRf+.
value for current output: See Table 2-6.
value for voltage output: See Table 2-4.
Query response range <newline>
range returns the present setting. Response data type is NR3.