Agilent Technologies B2902 Manuale di Servizio Pagina 57

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 326
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 56
Agilent B2900 SCPI Command Reference, Edition 2 2- 17
Subsystem Command Summary
Setting Source/Measure Unit
:SENS[c]:CURR[:DC]:RANG:AUTO:LLI
M range
:SENS[c]:RES:RANG:AUTO:LLIM rang
e
:SENS[c]:VOLT[:DC]:RANG:AUTO:LLI
M range
:SENS[c]:CURR[:DC]:RANG:AUTO:LLI
M? [DEFault|MINimum|MAXimum]
:SENS[c]:RES:RANG:AUTO:LLIM? [DE
Fault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:RANG:AUTO:LLI
M? [DEFault|MINimum|MAXimum]
Specifies the lower limit for the automatic
measurement ranging operation, and sets
the minimum measurement range which
provides the best resolution to measure
the specified value.
range=MINimum|MAXimum|DEFault|
minimum to maximum measurement
value, in A, , or V. See
“Measurement
Ranges” on page 2-27.
1.00E-06 A,
2 , or 0.2 V
:SENS[c]:CURR[:DC]:RANG:AUTO:MOD
E mode
:SENS[c]:VOLT[:DC]:RANG:AUTO:MOD
E mode
:SENS[c]:CURR[:DC]:RANG:AUTO:MOD
E?
:SENS[c]:VOLT[:DC]:RANG:AUTO:MOD
E?
Selects the operation mode of the
automatic measurement ranging. This
command setting is not effective if the
automatic ranging is disabled by the
:SENS:<CURR|VOLT>:RANG:AUTO
command.
mode=NORMal|RESolution|SPEed
NORM
:SENS[c]:CURR[:DC]:RANG:AUTO:THR
rate
:SENS[c]:VOLT[:DC]:RANG:AUTO:THR
rate
:SENS[c]:CURR[:DC]:RANG:AUTO:THR
? [DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:RANG:AUTO:THR
? [DEFault|MINimum|MAXimum]
Sets the threshold rate for the automatic
measurement ranging operation.
rate=MINimum|MAXimum|DEFault|
11
% to 100 %
90
Command Summary Reset setting
Vedere la pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 325 326

Commenti su questo manuale

Nessun commento