Agilent Technologies B2902 Manuale di Servizio Pagina 202

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 326
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 201
4- 80 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:THReshold
mode returns NORM, RES, or SPE, and indicates that the operation mode of
automatic measurement ranging. Response data type is CRD.
Example :SENS:CURR:RANG:AUTO:MODE SPE
:SENS2:CURR:DC:RANG:AUTO:MODE?
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AU
TO:THReshold
Sets the threshold rate for the automatic measurement ranging operation.
Syntax :SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:THReshold rate
:SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:THReshold? [rate]
For <CURRent[:DC]|VOLTage[:DC]>, specify CURRent[:DC] for current
measurement, or VOLTage[:DC] for voltage measurement.
Parameter rate value|MINimum|MAXimum|DEFault (default is 90 %).
Parameter data type is NRf+. Effective values of value are 11 %
to 100 %. Query does not support rate=value.
Query response rate <newline>
rate returns the present setting of the threshold rate for automatic measurement
ranging. If a parameter is specified, rate returns the value assigned to DEF, MIN, or
MAX. Response data type is NR3.
Example :SENS:CURR:RANG:AUTO:THR 60
:SENS2:CURR:DC:RANG:AUTO:THR?
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
RANGe:AUTO:ULIMit
Specifies the upper limit for the automatic measurement ranging operation, and sets
the maximum measurement range which provides the best resolution to measure the
specified value. This is effective for resistance measurements set to the AUTO mode
by the
:SENSe:RESistance:MODE command.
For current measurement and voltage measurement, the maximum measurement
range depends on the compliance range.
Vedere la pagina 201